Specifications
Gas: CO2
Sensor: INE20-CO2P-NCVSP
Performance Characteristics
Nominal Range: 0 to 5000 ppm
Long Term Output Drift: < ±250 ppm/year
Warm up time: 60 seconds @ 25 ºC
At least 30 min for full specification @ 25 °C
Response Time (T90): ≤ 60 seconds
Resolution: 25 ppm
Accuracy: as good as ±50 ppm*, from 0 to 2500 ppm range (ideal conditions)
as good as ±200 ppm*, from 2500 to 5000 ppm range (ideal conditions)
Operation Conditions
Temperature Range: -40 ºC to 60 ºC
Operating Humidity: 0 to 95% RH non-condensing
Storage Temperature: -40 ºC to 85 ºC
MTBF: ≥ 5 years
Average consumption: 80 mA
Note: The CO2 Sensor and the Methane (CH4) and Combustible Gas Sensor have high power requirements and cannot work together in the same Gases PRO Sensor Board. The user must choose one or the other, but not both