The Atlas Scientific industrial conductive probe is designed for inline conductivity measurements. From industrial chemical processing to swimming pool monitoring, this probe can be used for just about anything. The 3/4″ female NPT threading and integrated PT-1000 temperature probe makes for simple installation and operation.
Because conductivity probes don’t have any electrolyte that gets depleted, they only need to be calibrated during installation.
Specification
Reads | Conductivity |
Range | 0.07 − 50,000 μS/cm |
Accuracy | +/ – 2% |
Response time | 90% in 1s |
Temperature range °C | 0 − 110 °C |
Max pressure | 200 PSI |
Max depth | 60m (197 ft) |
Connector | Tinned leads |
Cable length | 3 meters (10′) |
Pipe fitting | 3/4″ threaded NPT |
Internal temperature sensor | Yes (PT-1000) |
Time before recalibration | ~10 years |
Life expectancy | ~10 years |